Digital, Bench-model Conductivity Meter – HI2030

Digital, Bench-model Conductivity Meter – HI2030

SKU: hi2030 Categories: ,

Description

The Hanna edge is thin and lightweight, measuring just ½” thick and weighing less than 9 ounces. edge has an incredibly wide viewing angle, 5.5″ LCD and a sensitive capacitive touch keypad edge measures pH, conductivity or dissolved oxygen through its unique digital electrodes. These digital electrodes are auto-recognized. Providing sensor type, calibration data and a serial number, and they connect to edge with an easy to plug-in 3.5mm connector. edge’s versatile design is equally at home in your hand, on a lab bench or mounted on a wall. edge simplifies measurement, configuration, calibration, diagnostics, logging and transferring data to a computer or a USB drive.
edge features Hanna’s exclusive CAL-CHECK to warn you if the electrode you are using is not clean or if your buffers are contaminated during calibration. We have also added to CAL-CHECK for sensors with matching pin: now it warns you if the pH bulb is cracked, and if the junction of the electrode is compromised.
The electrodes used with edge are nearly as advanced as the edge itself: featuring a built-in microchip that stores sensor type, ID, and calibration information that is automatically retrieved by edge once the electrode is plugged in.
Stored pH calibration information includes: calibrated buffers, date, time, offset and slope characteristics of the electrode. Conductivity calibration information includes: calibrated conductivity standards, date, time, and cell
constant of the sensor. Dissolved oxygen calibration information includes: standards used for calibration, date, time, altitude and salinity correction.
These digital electrodes also feature an easy to plug in 3.5 mm connector so you never have to worry about the right angle or aligning pin settings.

Logging: stores up to 1000 records: 200 records (log-on-demand and stability logging); 600 records interval logging
Connectivity: 1 USB port for storage; 1 micro USB port for charging and PC connectivity

EC:
EC TDS Salinity
Range: 0.00 to 29.99 µS/cm; 30.0 to 299.9µS/cm; 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 0.0 to 400.0 % NaCl;
300 to 2999 ?S/cm; 30.0 to 200.0 mS/cm; mg/L (ppm); 150 to 1499 mg/L (ppm); 0.01 to 42.00 PSU
up to 500.0 mS/cm (absolute EC)** 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; 0.0 to 80.0 g/L
up to 400.0 g/L (absolute TDS)**, with
0.80 conversion factor
———————————————————————————————————————————-
Resolution: 0.01 ?S/cm; 0.1 ?S/cm; 1 µS/cm; 0.01 mg/L (ppm); 0.1 mg/L (ppm);
0.01 mS/cm; 0.1 mS/cm 1 (ppm) 0.01 g/L; 0.1 g/L (0.8 TDS 0.1 % NaCl; 0.01 PSU; .01 g/L
factor)
———————————————————————————————————————————-
Accuracy (@25°C): ±1% of reading ±(0.5 µS or ±1% of reading ±(0.03 ppm or 1 digit, ±1% of reading
1 digit, whichever is greater) whichever is greater)
———————————————————————————————————————————-
Calibration: 1 point offset calibration (0.00 µS/cm
in air); 1 point slope calibration in 1 point with HI 7037L 100% NaCl
EC standard 84 µS/cm, 1413 µS/cm, 5.00 through EC calibration sea water standard (other scales
mS/cm, 12.88 mS/cm, 80.0 mS/cm and through EC calibration)
118.8 mS/cm
———————————————————————————————————————————-
Temperature Compensation: noTC, ATC (0.0 to 100.0°C; 32.0 to 212.°F
———————————————————————————————————————————-
TDS Factor: 0.40 to 0.80

Temperature:
Range: -20.0 to 120.0°C; -4.0 to 248.0°F
Resolution: 0.1°C; 0.1°F
Accuracy: ±0.2°C; ±0.4°F

INCLUDED: edge EC Kit, 115V, including edge meter, HI 763100 digital 4-ring conductivity electrode, bench cradle, wall crade, electrode holder, USB cable, 5 Vdc power adapter, instruction manual and solution starting kit. Request pricing on buffers or storage solutions
Depending on inventory and destination, this item ships from Richmond Hill, ON or Laval, QC.